Advanced Pendulum Impact System
The Pendulum Tester uses microprocessor technology to determine the energy required to break or rupture specimens in plastic or ceramics. The apparatus can test Izod , Charpy or Tension Impact , as well as carry out small diameter pipe testing and component Impact & Puncture tests.
Testing methods available:
Izod Test
Notched or Un-notched specimens can be tested using the Izod method. Complies with International Test Standards ASTM D256, ASTM D4812, and ISO 180. Impact hammer energies are available from 0.5 J up to 25 J
Charpy Test
Complies with International Test Standards ISO 179, ASTM D256, ASTM D6110, DIN 53453, DIN 53753, and BS 7413. Impact hammer energies are available from 0.5 J up to 50 J
Tension Impact
Complies to International Test Standard ISO 8256 Method A. Impact hammer energies are available from 0.5 J up to 50 J.
Pipe Testing
Sample diameters up to 25 mm can be tested with hammer energies of 7.5 – 15 J. The test specimen is supported as a horizontal beam similar to Charpy test conditions and complies with International Test Standards ISO 7628 and ISO 9854.