Specification
Product Specification | Description |
Conductivity Range | .. to 19.99 uS/cm 19.9 uS/cm to 199.9 uS/cm 199.9 uS/cm to 1999 µS/cm 2.00 mS/cm to 19.99 mS/cm 20.0 mS/cm to 199.9 mS/cm |
Resolution & Accuracy | 0.05% full scale & ±1% full scale + 1 LSD |
Calibration Points | 4 (1 per range) auto; 5 (1 per range) manual |
Temperature Range | 0 to 80.0 °C (32 to 176 °F) |
Resolution & Accuracy | 0.1 °C (0.1 °F) & ±0.5 °C (±0.9 °F) |
Temperature Compensation | Automatic / Manual (0 to 80 °C) |
Cell Constant | 0.1, 1.0, 10.0 (selectable) |
Temperature Coefficient | 0.0% to 3.0% |
Temperature Normalisation | 20.0 to 25.0 °C |
No. of Calibration Points | Up to 5 max. (1 per range) |
Special Functions | Auto Power-off after 17 minutes; Hold & Self-diagnostic messages |
Power | 4 ‘AAA’ x 1.4V batteries; >100 hours |